Z-0072-2017Class II

ADVIA Centaur XPT Immunoassay System

Siemens Healthcare Diagnostics Inc. / initiated 2016-04-28 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA Centaur XPT Immunoassay System on . FDA classified it as Class II and gives the reason as "The ADVIA Centaur XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0072-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Centaur XPT Immunoassay System
Product code
JLW Radioimmunoassay, Thyroid-Stimulating Hormone
Reason for recall
The ADVIA Centaur XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Siemens sent an Urgent Medical Device Correction letter dated April 28, 2016, to all affected customers. The letter advised customers not to release results generated by the system if quality control results are not within acceptable ranges. Customers were advised to perform the following steps to check the Setting for the Daily Maintenance Task frequency on your instrument: *Open the Maintenance Tab, * Select Daily Maintenance task and * Verify that the Frequency Column is set to "Daily at..". The Daily Maintenance Task frequency setting can only be reset by a Siemens Customer Service engineer. If your instrument is set to "As Needed": Daily Maintenance Task Frequency Setting May be Incorrect. Customers should contact their local Technical Support Center to schedule a visit by your Siemens Customer Service engineer. Customers should also perform the daily maintenance by selecting the Daily Maintenance Task and then selecting Perform on the right bar. In addition, all customers were asked to complete and return the Field Correction Effectiveness Check within 30 days. For questions regarding this recall call 914-631-8000.
Quantity in commerce
292 units Total (3 domestically & 289 internationally)
Distribution
Worldwide Distribution - US including MD, MN, & MS; **Internationally** AE, AT, AU, BE, BR, CA, CN, DE, DK, EG, ES, FI, FR, GB, GP, HU, IN, IT, JO, JP, KR, MY, NO, NZ, PH, PT, QA, SA, SE, SG, TR & VN
Product codes and lots
None

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0072-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0072-2017Class IITerminatedVoluntary: Firm initiated