Z-1703-2016Class II

WHIN Infusion Set - 90 Huber Needle and Wing (22 ga/1in). Intravenous delivery through intravascular ports...

B Braun Medical Inc / initiated 2016-04-26 / Terminated / root cause: design

B. Braun Medical, Inc. began a recall of WHIN Infusion Set - 90 Huber Needle and Wing (22 ga/1in). Intravenous delivery through intravascular ports, 20 units per case, Catalog Number 471734. Usage... on . FDA classified it as Class II and gives the reason as "Potential for the tip of the Huber needle on the WHIN Infusion Set to protrude from the needle guard". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1703-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B Braun Medical Inc
Product
WHIN Infusion Set - 90 Huber Needle and Wing (22 ga/1in). Intravenous delivery through intravascular ports, 20 units per case, Catalog Number 471734. Usage: Infusion of solution/medication into implanted intravascular ports.
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Potential for the tip of the Huber needle on the WHIN Infusion Set to protrude from the needle guard.
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm, B. Braun Medical Inc., sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated 4/26/2016 to customers. The letter described the product, problem and actions to be taken. The customers were instructed to evaluate their current inventory for the affected lots; fill out the Product Removal Acknowledgement form and record the total number of units found, and return the form to B. Braun Medical Inc. Quality Assurance department by faxing the form to 610-849-1197 or email to PA_Quality [email protected] within two weeks of receipt. A BBMI Customer Support Representative will contact the customer to provide instructions for handling the affected product and arrange for return to BBMI. Should you have any questions or concerns regarding the attached information, please contact our Customer Support Department at (800) 227-2862.
Quantity in commerce
5,980 units [Lot # 0061482060 (3,340 units); Lot # 0061486821 (2,640 units)]
Distribution
Worldwide distribution: US Distribution to states of: IN, NV, MA, IL, AR, TX, OH, CA, GA, FL, MS, NJ, AL, PA, TN, KY, VA, SC, and country of: Canada.
Product codes and lots
Catalog Number 471734, Lot Numbers: 0061482060 and 0061486821, expiration dates :1/31/2021 and 02/28/2021, respectively.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1703-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1703-2016Class IITerminatedVoluntary: Firm initiated