510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1837-2016Class II

SEAL,IS4000 PORTS,12MM&STAPLER,BOX of 10; Model numbers: 470380-04, 470380-05, 470380-06; Part number:470380...

Intuitive Surgical, Inc. / initiated 2016-05-09 / Terminated / root cause: manufacturing process

Intuitive Surgical, Inc. began a recall of SEAL,IS4000 PORTS,12MM&STAPLER,BOX of 10; Model numbers: 470380-04, 470380-05, 470380-06; Part number:470380. The da Vinci Xi Surgical System cannula seal is... on . FDA classified it as Class II and gives the reason as "Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula lumen when insufflation is connected during a surgical...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1837-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
SEAL,IS4000 PORTS,12MM&STAPLER,BOX of 10; Model numbers: 470380-04, 470380-05, 470380-06; Part number:470380. The da Vinci Xi Surgical System cannula seal is used with the cannula accessory and provides sterile disposable cap for each cannula. When attached to the cannula, the seal maintains insufflation when the cannula is empty and when the instrument is inserted during a surgical procedure.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula lumen when insufflation is connected during a surgical procedure.
Root cause tag
manufacturing process
FDA root cause
Manufacturing material removal
Action
Upon discovery of the issue, Intuitive sent Urgent Product Safety Notices to all customers by e-mail to initially alert them of the situation. This will be followed by a formal letter that is scheduled to arrive on May 23, 2016. Both notification request that users Do NOT use the da Vinci Xi 5 mm8 mm Universal Cannula Seal or the da Vinci Xi 12 mm & Stapler Universal Cannula Seal for insufflation, or open the stopcock during intracardiac procedures.
Quantity in commerce
46,191 boxes total - both sizes
Distribution
Worldwide distribution. US, Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1837-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1836-2016Class IITerminatedVoluntary: Firm initiated
Z-1837-2016Class IITerminatedVoluntary: Firm initiated