Biograph Sys 16-3R to 16-4R upgrade, MATERIAL NUMBER 10525581 The Siemens Biograph TruePoint systems are...
Siemens Medical Solutions USA Inc. began a recall of Biograph Sys 16-3R to 16-4R upgrade, MATERIAL NUMBER 10525581 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron... on . FDA classified it as Class II and gives the reason as "Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2017-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Biograph Sys 16-3R to 16-4R upgrade, MATERIAL NUMBER 10525581 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens sent a Customer Safety Advisory Letter on April 22, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Siemens representative at the contact numbers provided in the letter.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide Distribution - US Nationwide
- Product codes and lots
- Serial Numbers: 72418,72428
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2005-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2006-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2007-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2008-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2009-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2010-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2011-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2012-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2013-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2014-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2015-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2016-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2017-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2018-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2019-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2020-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2021-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2022-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2023-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2024-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2025-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2026-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2027-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2028-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2029-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2030-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2031-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2032-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2033-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2034-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2035-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2036-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2037-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.