Z-1997-2016Class II

BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10507786

Siemens Medical Solutions USA, Inc. / initiated 2016-04-25 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA Inc. began a recall of BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10507786 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron... on . FDA classified it as Class II and gives the reason as "The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1997-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10507786 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Siemens sent a Customer Safety Advisory Letter to all affected customers. The letter identified the affected product, problem and actions to be taken. Siemens service organization will be contacting customers to schedule the update as per the attached Update Instruction and will be provided the replacement bracket to resolve the issue. For questions contact your Siemens representative.
Quantity in commerce
1 unit
Distribution
Worldwide Distribution - US Nationwide and the countries of China and Malaysia.
Product codes and lots
Serial Number 21139

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1997-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1997-2016Class IITerminatedVoluntary: Firm initiated
Z-1998-2016Class IITerminatedVoluntary: Firm initiated
Z-1999-2016Class IITerminatedVoluntary: Firm initiated
Z-2000-2016Class IITerminatedVoluntary: Firm initiated