Z-2248-2016Class II

OT1000 Series Orthopedic Surgical Tables

STERIS Corporation / initiated 2016-05-23 / Terminated

Steris Corporation began a recall of OT1000 Series Orthopedic Surgical Tables on . FDA classified it as Class II and gives the reason as "STERIS has identified that the pivot pin assembly located on the table top may become loose preventing the table from articulating properly". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2248-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
OT1000 Series Orthopedic Surgical Tables
Product code
JEA Table, Surgical With Orthopedic Accessories, Ac-Powered
Reason for recall
STERIS has identified that the pivot pin assembly located on the table top may become loose preventing the table from articulating properly.
FDA root cause
Other
Action
Recall Notification Letters to the affected Consignees was sent on May 23, 2016 via FedEx with tracking numbers for delivery confirmation. STERIS Field Service Technicians will visit each domestic Customer location to perform the inspection and reinstallation of the pivot pin assemblies. For international direct accounts who have further distributed the surgical table, Steris asked to identify the Customers, and to notify them of this product field correction, and that not to operate the tables until the inspection has been done of the pivot pin assembly on all affected tables.
Quantity in commerce
148
Distribution
AL;AK, AZ; AR; CA; CO; CT; DC; FL; GA; IL; IN; IA; NY; NJ; NC; OH; OK; OR; PA; PR; SC; TX; UT and VA. Foreign accounts: Canada; France; Korea; Panama; Qatar and Thailand.
Product codes and lots
OT1000, OT1100, and OT1200 Orthopedic Surgical Table