Z-2087-2016Class II
Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge...
Siemens Healthcare Diagnostics Inc. / initiated 2016-05-11 / Terminated
Siemens Healthcare Diagnostics Inc. began a recall of Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High... on . FDA classified it as Class II and gives the reason as "Premature bucket failures". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2087-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High Throughput Centrifuge Module (11171747) FlexLab and Aptio Centrifuge Module (10628157 or 11266581): Used to separate substances according to their densities.
- Product code
- LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use
- Reason for recall
- Premature bucket failures
- FDA root cause
- Under Investigation by firm
- Action
- Siemens mailed to customers an Urgent Medical Device Correction (UMDC#LAI16-02.A.US) Urgent Field Safety Notification letter informing customers of the problem with recommendations to follow the existing maintenance instructions in the product operator manuals as well as a new requirement to spin at cycle speeds not exceeding 3500 rpm.
- Quantity in commerce
- 487
- Distribution
- Nationwide: AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, WV.
- Product codes and lots
- All serial numbers are affected.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2087-2016 | Class II | Terminated | Voluntary: Firm initiated |