Z-2087-2016Class II

Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge...

Siemens Healthcare Diagnostics Inc. / initiated 2016-05-11 / Terminated

Siemens Healthcare Diagnostics Inc. began a recall of Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High... on . FDA classified it as Class II and gives the reason as "Premature bucket failures". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2087-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High Throughput Centrifuge Module (11171747) FlexLab and Aptio Centrifuge Module (10628157 or 11266581): Used to separate substances according to their densities.
Product code
LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use
Reason for recall
Premature bucket failures
FDA root cause
Under Investigation by firm
Action
Siemens mailed to customers an Urgent Medical Device Correction (UMDC#LAI16-02.A.US) Urgent Field Safety Notification letter informing customers of the problem with recommendations to follow the existing maintenance instructions in the product operator manuals as well as a new requirement to spin at cycle speeds not exceeding 3500 rpm.
Quantity in commerce
487
Distribution
Nationwide: AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, WV.
Product codes and lots
All serial numbers are affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2087-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2087-2016Class IITerminatedVoluntary: Firm initiated