Z-2081-2016Class II
CENTURION Vision System
Alcon Laboratories / initiated 2016-05-24 / Terminated / root cause: manufacturing process
Alcon Research, Ltd. began a recall of CENTURION Vision System on . FDA classified it as Class II and gives the reason as "The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2081-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Laboratories
- Product
- CENTURION Vision System
- Product code
- HQC Unit, Phacofragmentation
- Reason for recall
- The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Alcon has already notified the affected customer verbally on May 24, 2016, and via written correspondence on June 1, 2016. The device was corrected on May 26, 2016, and the Power Entry Switch that was removed from the customers CENTURION Vision System did not exhibit a grounding pin susceptible to detachment. For further questions, please call (817) 293-0450.
- Quantity in commerce
- 1 unit
- Distribution
- US Distribution to Illinois
- Product codes and lots
- Serial Number 1601444801X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2081-2016 | Class II | Terminated | Voluntary: Firm initiated |