510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2081-2016Class II

CENTURION Vision System

Alcon Laboratories / initiated 2016-05-24 / Terminated / root cause: manufacturing process

Alcon Research, Ltd. began a recall of CENTURION Vision System on . FDA classified it as Class II and gives the reason as "The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2081-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alcon Laboratories
Product
CENTURION Vision System
Product code
HQC Unit, Phacofragmentation
Reason for recall
The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Alcon has already notified the affected customer verbally on May 24, 2016, and via written correspondence on June 1, 2016. The device was corrected on May 26, 2016, and the Power Entry Switch that was removed from the customers CENTURION Vision System did not exhibit a grounding pin susceptible to detachment. For further questions, please call (817) 293-0450.
Quantity in commerce
1 unit
Distribution
US Distribution to Illinois
Product codes and lots
Serial Number 1601444801X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2081-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2081-2016Class IITerminatedVoluntary: Firm initiated