Covidien Cytosponge Cell Collection Device. Item code CYTO-101-01. Gastroenterology: -.Cytosponge Cell...
Covidien, LLC began a recall of Covidien Cytosponge Cell Collection Device. Item code CYTO-101-01. Gastroenterology: -.Cytosponge Cell Collection Device is indicated for use in the collection... on . FDA classified it as Class II and gives the reason as "Medtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal string during withdrawal from the patient's esophagus". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2123-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Cytosponge Cell Collection Device. Item code CYTO-101-01. Gastroenterology: -.Cytosponge Cell Collection Device is indicated for use in the collection and retrieval of surface cells in the esophagus. This device is currently only being used in clinical studies.
- Product code
- EOX Esophagoscope (Flexible Or Rigid)
- Reason for recall
- Medtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal string during withdrawal from the patient's esophagus.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, Medtronic, sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 6, 2016 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to quarantine and discontinue use; return affected product to Medtronic/Covidien, Quality Department, 540 Oakmead Parkway, Sunnyvale CA 94085; complete and return the RECALLED PRODUCT RETURN FORM by email to Customer Service feedback.customerservice@Covidien or fax to (800)-895-6140 if product purchased directly from Medtronic; or by email to Quality Compliance [email protected] or fax to (203) 492-7719 if Product purchased through distributor or you have "no inventory"; and to promptly alert any other person or facilities if they have further distributed the product. All affected lots must be returned. If you have any questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at (800) 882-5878.
- Quantity in commerce
- 620 total units (260 US; 360 OUS)
- Distribution
- Worldwide Distribution-US Distribution to TN, NC, IL, CA, MN, OR and NY; and countries of: United Kingdom and Iran.
- Product codes and lots
- Item Code: CYTO-101-01. Lot numbers: F2500566X, F2500628X, F2500202X, F2500351X.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2122-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2123-2016 | Class II | Terminated | Voluntary: Firm initiated |