enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and...
Ortho-Clinical Diagnostics began a recall of enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; IVD on . FDA classified it as Class III and gives the reason as "If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2143-2016
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; IVD
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends the incorrect text to the LIS. The IFU states the interpretation term for these numerical result values is Retest?. The difference in terminology exists for US Markets only.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Ortho Clinical Diagnostics sent an "IMPORTANT PRODUCT CORRECTION NOTIFICATION" Customer Letter dated May 16, 2016, to all affected customers in the US with enGen Systems via FedEx or US Priority Mail (for Post Office Box addresses) or by e-mail to subscribers of ORTHO PLUS e-Communications to advise them of the issue and the required actions . Required Actions: 1. Use the numerical test results to assess actions to be taken for these assays, per the IFU. 2. Post this notification by your enGen System or with your user documentation. 3. Complete and return the Confirmation of Receipt form by 25 May 2016. with the Instructions for Use (IFU) directions). If you have any questions regarding this notification, please call the Ortho Technical Solutions Center at any time at 1-800-421-3311.
- Quantity in commerce
- US: 30 units
- Distribution
- Nationwide
- Product codes and lots
- Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; Serial Numbers/J Numbers: JIM341866, JIM341820, JIM341679, JIM341352, JIM343421, JIM341174, JIM336745, JIM340667, JIM338283, JIM-341592, JIM336957, JIM341996, JIM342334, JIM343694, JIM336855, JIM334236, JIM341415, JIM341392, JIM336481, JIM340934, JIM336167, JIM342249, JIM338290, JIM341750, JIM341426, JIM334986, JIM334527, JIM340024, JIM341021, JIM340658
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2143-2016 | Class III | Terminated | Voluntary: Firm initiated |