VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No...

Ortho-Clinical Diagnostics, Inc. / initiated 2016-05-18 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No. 10758750009114 or 20758750009111, IVD -- Sales Unit: 300... on . FDA classified it as Class II and gives the reason as "Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be generated when using VITROS Ca Slides, GEN29, and the VITROS 250/350/950/5,1 FS, 4600 Chemistry System or the VITROS 5600...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2273-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No. 10758750009114 or 20758750009111, IVD -- Sales Unit: 300 slides/pack Product Usage: For in vitro diagnostic use only. VITROS Ca Slides quantitatively measure calcium (Ca) concentration in serum, plasma, and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone disorders, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Product code
CJY Azo Dye, Calcium
Reason for recall
Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be generated when using VITROS Ca Slides, GEN29, and the VITROS 250/350/950/5,1 FS, 4600 Chemistry System or the VITROS 5600 Integrated System. An increase in negatively biased calcium results complaints prompted the investigation that identified this product issue.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION letter dated May 18, 2016 to affected customers. . The letter identified the affected product, problem and actions to be taken. The letter instructed to discontinue use of the affected lot(s) and discard all remaining inventory. Customers were asked to complete and return the Confirmation of Receipt form. For questions contact Customer Technical Services at 1-800-421-3311.
Quantity in commerce
US: 27,625 sales units; Foreign: 4757 sales units
Distribution
Worldwide Distribution - US Nationwide in the countries of Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, Poland and Venezuela.
Product codes and lots
Lots #s: 0329-0528-8193 (exp. 05/01/2017), 0329-0528-8892 (exp. 06/01/2017), 0329-0528-8192 (exp. 05/01/2017), 0329-0528-8902 (exp. 06/01/2017), 0329-0528-8226 (exp. 05/01/2017), 0329-0528-8928 (exp. 07/01/2017),  0329-0528-8227 (exp. 05/01/2017), 0329-0528-9000 (exp. 07/01/2017), 0329-0528-8359 (exp. 05/01/2017), 0329-0528-8927 (exp. 07/01/2017), 0329-0528-8361 (exp. 05/01/2017), 0329-0528-8965 (exp. 07/01/2017), 0329-0528-8360 (exp. 05/01/2017), 0329-0528-8966 (exp. 07/01/2017),  0329-0528-8324 (exp. 05/01/2017), 0329-0528-9167 (exp. 07/01/2017), 0329-0528-8323 (exp. 05/01/2017), 0329-0528-9171 (exp. 07/01/2017), 0329-0528-8325 (exp. 05/01/2017), 0329-0528-9196 (exp. 07/01/2017), 0329-0528-6999 (exp. 03/01/2017), 0329-0528-9172 (exp. 07/01/2017), 0329-0528-7000 (exp. 03/01/2017), 0329-0528-9177 (exp. 07/01/2017), 0329-0528-6998 (exp. 03/01/2017), 0329-0528-9265 (exp. 07/01/2017), 0329-0528-8903 (exp. 06/01/2017), 0329-0528-9176 (exp. 07/01/2017), 0329-0528-8901 (exp. 06/01/2017), 0329-0528-9264 (exp. 07/01/2017), 0329-0528-8893 (exp. 06/01/2017), 0329-0528-9368 (exp. 07/01/2017), 0329-0528-8891 (exp. 06/01/2017), 0329-0528-9369 (exp. 07/01/2017).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2273-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2273-2016Class IITerminatedVoluntary: Firm initiated