Z-2043-2016Class II

Fujifilm Synapse PACS software version 4.4.000, Fujifilm Synapse PACS software version 4.4.001, Fujifilm...

FUJIFILM Healthcare Americas Corporation / initiated 2016-05-10 / Terminated / root cause: software

Fujifilm Medical Systems U.S.A., Inc. began a recall of Fujifilm Synapse PACS software version 4.4.000, Fujifilm Synapse PACS software version 4.4.001, Fujifilm Synapse PACS software version 4.4.004, Fujifilm... on . FDA classified it as Class II and gives the reason as "Synapse cannot display image files, DICOM SR files, and/or Annotation files. The "Image Not Loaded" message displays instead". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2043-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Fujifilm Synapse PACS software version 4.4.000, Fujifilm Synapse PACS software version 4.4.001, Fujifilm Synapse PACS software version 4.4.004, Fujifilm Synapse PACS software version 4.4.010 and Fujifilm Synapse PACS software version 4.4.020 FUJIFILM Synapse Workstation Software (a Picture Archiving And Communications System) is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. Also intended for installation on an off-the-shelf PC networked with Fuji Synapse PACS.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Synapse cannot display image files, DICOM SR files, and/or Annotation files. The "Image Not Loaded" message displays instead
Root cause tag
software
FDA root cause
Software design
Action
The firm, FujiFilm, sent an "MEDICAL DEVICE CORRECTION" letter dated May 10, 2016 via Federal Express to its customers. The letter describes the product, problem and actions to be taken.The customers were instructed to: Extend the Ad hoc deletion retention period or discontinue use of Ad Hoc Deletion, and complete and submit the attached Customer Response Form to Fujifilm via fax to (203)-251-7863. Affected systems will be corrected via on-site visits by FMSU Customer Service Engineers upgrading to Version 4.4.100. Questions or concerns, contact the FUJI FILM Customer Support Center at 1- 888-FUJIMED Option #1 .
Quantity in commerce
27 units
Distribution
US Distribution to states of: CA, FL, KS, NE, NY, and PA.
Product codes and lots
Software version: 4.4.000, 4.4.001, 4.4.004, 4.4.010, 4.4.020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2043-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2043-2016Class IITerminatedVoluntary: Firm initiated