510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2210-2016Class II

Preamendments device-Sarns Dual Cooler-Heater. The Sarns Dual Cooler-Heater is a self-contained unit for...

Terumo Cardiovascular Systems Corporation / initiated 2016-06-08 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Preamendments device-Sarns Dual Cooler-Heater. The Sarns Dual Cooler-Heater is a self-contained unit for supplying cold and warm water to heat exchangers... on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Product Code 205) referenced in its...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2210-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Preamendments device-Sarns Dual Cooler-Heater. The Sarns Dual Cooler-Heater is a self-contained unit for supplying cold and warm water to heat exchangers during open heart surgery. It supplies water at a rate of up to 6.5 gallons per minute (25 liters per minute) at zero head, and will provide a maximum pressure output of 11 pounds per square inch (76,000 pascals). The Sarns TM Dual Cooler-Heater can be used with heat exchangers capable of accepting these flows and pressures. Product Usage: The Sarns Dual Cooler-Heater is indicated for use as a water supply system to a blood heat exchanger in open-heart surgery.
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Terumo Cardiovascular Systems (Terumo CVS) initiated a voluntary recall on 06/08/2016, by mailing certifed letters to customers. Terumo CVS is notifying all users of its heater-cooler devices that the distributor, Intercon Chemical Company, no longer sells the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions. However, the identical sanitizer is available from other sources. This notice applies to the following Terumo CVS heater-cooler devices: Sarns" TCM and TCM II Cooling and Heating Systems Sarns" Dual Cooler-Heaters HX2" Temperature Management Systems Terumo CVS will provide users with updated user instructions regarding the 10% Sodium Hypochlorite sanitization solution. Instructions will be provided under separate cover, when available; updated manual and cleaning guide will be completed in December 2016. Customers are asked to do the following: Review this notice and ensure that all users receive notice of this issue. 2. Return the attached Customer Response Form acknowledging receipt of this notice. 3. Source a replacement for the Interchlor sanitizer. See information above or call Terumo CVS Technical Support for assistance. If you have questions about this notification, please contact your Terumo representative or Terumo CVS Technical Support: 800.441.3220, extension 6932 Technical Support Hours: Monday Friday 8 a.m. 6 p.m. ET (24/7 Emergency Hours)
Quantity in commerce
6412 units
Distribution
Domestic: AK, AL, AR, AZ, CA , CO, CT, DC , DE , FL, GA, HI, IA, ID, IL, IN, KS, KY,LA , MA, MD, ME, MI , MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV , NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign: Argentina, Australia, Belgium,Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, England, Finland, France, Germany Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Malaysia Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Portugal, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates , Uruguay , Venezuela, Vietnam
Product codes and lots
The Sarns Dual Cooler-Heater ; Dual Cooler Heater 115V/60Hz (P/N 11160); Dual Cooler Heater 220V/50Hz (P/N 4332); Dual Cooler Heater 1 OOV/60Hz (P/N 14114); Dual Cooler Heater 1 OOV/50Hz (PIN 13460)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2210-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2207-2016Class IITerminatedVoluntary: Firm initiated
Z-2208-2016Class IITerminatedVoluntary: Firm initiated
Z-2209-2016Class IITerminatedVoluntary: Firm initiated
Z-2210-2016Class IITerminatedVoluntary: Firm initiated