510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2108-2016Class II

Access Free T3 Assay, Ref No. A13422 The Access Free T3 assay is a paramagnetic particle, chemiluminescent...

Siemens Healthcare Diagnostics Inc. / initiated 2016-06-08 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of Access Free T3 Assay, Ref No. A13422 The Access Free T3 assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of... on . FDA classified it as Class II and gives the reason as "Demonstrate an upward shift in patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2108-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Free T3 Assay, Ref No. A13422 The Access Free T3 assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of free triiodothyronine levels in human serum and plasma using the Access Immunoassay Systems.
Product code
CDP Radioimmunoassay, Total Triiodothyronine
Reason for recall
Demonstrate an upward shift in patient results.
Root cause tag
design
FDA root cause
Device Design
Action
Beckman Coulter sent an " Urgent Medical Device Recall" letter dated June 8, 2016, to all affected customers. Beckman Coulter has determined through customer feedback and internal testing that the Access Free T3 Reagent lots listed above demonstrate an upward shift in patient results. The preliminary results indicate this upward shift may be related to a June 2015 formulation design change that was introduced to improve the Access Free T3 open reagent pack stability. The upward shift in patient test results, therefore, is expected to be maintained for all future lots. The letter informs the customers of the impact and the actions to be taken. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with any questions are instructed to contact Customer Technical Support Center at http:www.beckmancoulter.com, call (800) 854-3633 in US and Canada. Customers outside US and Canada, are instructed to contact local Beckman Coulter representative.
Quantity in commerce
205,744 kits total (13,358 kits in US)
Distribution
Worldwide Distribution - US Distribution and to the countries of : Algeria, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Djibouti, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Ghana, Gibraltar, Greece, Guam, Honduras, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lithuania, Macao, Malaysia, Mayotte, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam, Zambia, Zimbabwe
Product codes and lots
524087 526807 528874 529913 530102 531647 532473 534080 534081 534336 535481 570090 570116 570222 570236 570239 570245 570246 622173 622174 671016 671025 671033

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2108-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2108-2016Class IITerminatedVoluntary: Firm initiated