Screwdriver Blade, Handle, Drill Bit, Part #313.945 Low Profile Neuro Screwdriver Blade/MR Safe/MQC/32mm...
Synthes (USA) Products LLC began a recall of Screwdriver Blade, Handle, Drill Bit, Part #313.945 Low Profile Neuro Screwdriver Blade/MR Safe/MQC/32mm Short Lot # 5053859 Product Usage: The DePuy Synthes... on . FDA classified it as Class II and gives the reason as "DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2189-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Screwdriver Blade, Handle, Drill Bit, Part #313.945 Low Profile Neuro Screwdriver Blade/MR Safe/MQC/32mm Short Lot # 5053859 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.
- Product code
- HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
- Reason for recall
- DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- DePuy Synthes mailed an URGENT MEDICAL DEVICE FIELD SAFETY NOTIFICATION letter to their customers (Hospital Personnel, including Imaging Department Personnel and DePuy Synthes Colleague) to inform them of the recall and actions to take. The letter identified the affected, problem and actions to be taken. The letter instructed customers not to use the affected Instruments which are labeled or etched MR Safe and MR Conditional during MRI scans. Customers were asked to review the information contained in the Field Safety Notification letter and complete the Verification Section located on page 3 of this notification. For questions call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
- Quantity in commerce
- 50 units
- Distribution
- US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.
- Product codes and lots
- Part Number: 313.945, Low Profile Neuro Screwdriver Blade/MR Safe/MQC/32mm Short, Lot # 5053859, mfg. 7/28/2005, No Expiration Date product provided Non Sterile
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2189-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2190-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2191-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2192-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2193-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2194-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2195-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2196-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2197-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2198-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2199-2016 | Class II | Terminated | Voluntary: Firm initiated |