510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2194-2016Class II

Part # 313.211, 1.3mm Stardrive Screwdriver Bld Slf-Retain Shrt/42mm MQC, lot # 4341031, 4363779, 4427697...

Synthes (USA) Products LLC / initiated 2016-05-31 / Terminated / root cause: labeling and UDI

Synthes (USA) Products LLC began a recall of Part # 313.211, 1.3mm Stardrive Screwdriver Bld Slf-Retain Shrt/42mm MQC, lot # 4341031, 4363779, 4427697, mfg. 2/28/2002-6/6/2002 Product Usage: The DePuy... on . FDA classified it as Class II and gives the reason as "DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2194-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Part # 313.211, 1.3mm Stardrive Screwdriver Bld Slf-Retain Shrt/42mm MQC, lot # 4341031, 4363779, 4427697, mfg. 2/28/2002-6/6/2002 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.
Product code
HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
Reason for recall
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
DePuy Synthes mailed an URGENT MEDICAL DEVICE FIELD SAFETY NOTIFICATION letter to their customers (Hospital Personnel, including Imaging Department Personnel and DePuy Synthes Colleague) to inform them of the recall and actions to take. The letter identified the affected, problem and actions to be taken. The letter instructed customers not to use the affected Instruments which are labeled or etched MR Safe and MR Conditional during MRI scans. Customers were asked to review the information contained in the Field Safety Notification letter and complete the Verification Section located on page 3 of this notification. For questions call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
Quantity in commerce
75 units
Distribution
US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.
Product codes and lots
Part # 313.211, 1.3mm Stardrive Screwdriver Bld Slf-Retain Shrt/42mm MQC, lot # 4341031, 4363779, 4427697, mfg. 2/28/2002-6/6/2002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2194-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2189-2016Class IITerminatedVoluntary: Firm initiated
Z-2190-2016Class IITerminatedVoluntary: Firm initiated
Z-2191-2016Class IITerminatedVoluntary: Firm initiated
Z-2192-2016Class IITerminatedVoluntary: Firm initiated
Z-2193-2016Class IITerminatedVoluntary: Firm initiated
Z-2194-2016Class IITerminatedVoluntary: Firm initiated
Z-2195-2016Class IITerminatedVoluntary: Firm initiated
Z-2196-2016Class IITerminatedVoluntary: Firm initiated
Z-2197-2016Class IITerminatedVoluntary: Firm initiated
Z-2198-2016Class IITerminatedVoluntary: Firm initiated
Z-2199-2016Class IITerminatedVoluntary: Firm initiated