Z-2107-2016Class II

SYNGO Breast Care, visualization and image enhancement tools to aid radiologist in the review of digital...

Siemens Medical Solutions USA, Inc. / initiated 2016-06-06 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SYNGO Breast Care, visualization and image enhancement tools to aid radiologist in the review of digital Mammography images and tomosynthesis datasets on . FDA classified it as Class II and gives the reason as "Software issues. Siemens is voluntarily initiating a recall after they became aware of the following system behavior: 1) At times, the view and laterality marker is overlaid by patient demographic information, thus...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2107-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SYNGO Breast Care, visualization and image enhancement tools to aid radiologist in the review of digital Mammography images and tomosynthesis datasets.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Software issues. Siemens is voluntarily initiating a recall after they became aware of the following system behavior: 1) At times, the view and laterality marker is overlaid by patient demographic information, thus becoming difficult to read. 2) The facility name and address are not shown per default (in the current software versions this information is displayed only when the reader chooses an appropriate image text display mode).
Root cause tag
software
FDA root cause
Other
Action
Siemens mailed customers a letter to inform them about improvements that were made to the SYNGO Breast Care reading software to correct certain behaviors that were being seen during use. Customers were asked to upgrade their systems to this new version.
Quantity in commerce
29 units
Distribution
Distributed to: CA,NY,TX,OH,CO,NY,TX,CA,NE,NJ,TX,FL,IL,TX,MO,CA,PA,FL,NJ,MO,ND
Product codes and lots
Device Model # 10496180  Serial Numbers: 221176,130159,130773,100641,100642,100643,100644,100645,100811,100812,100813,100814,221129,130326,221192,130201,220930,130438,102113,130635,130377,130125,130546,130322,101309,101503,220743,102013,130575

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2107-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2107-2016Class IITerminatedVoluntary: Firm initiated