510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2118-2016Class II

Ingenuity Core 128-Computed Tomography X-ray system

Philips North America LLC / initiated 2016-04-01 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to... on . FDA classified it as Class II and gives the reason as "Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2118-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes..
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
Philips Medical Systems (Cleveland) Inc. sent an Urgent Field Safety Notice letter to affected customers , via certified letter with receipts returned for tracking purposes. The letter identified the affected product, problem and actions to be taken. Philips Healthcare is implementing software update version 3.5.5 to correct the problem. For questions contact your local Philips representative or local Philips Healthcare office.
Quantity in commerce
19
Distribution
US Nationwide Distribution in the states of CA, FL, ID, KS, MI, NY,OH,TX
Product codes and lots
Ingenuity Core128:System Code #728323, Serial numbers: 320001, 320016, 320076, 320077, 320080, 320085, 320086, 320087, 320088, 320089, 320091, 320113, 336001, 336002, 336003, 336006, 336007, 336008, 336009, 336010, 336014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2118-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2116-2016Class IITerminatedVoluntary: Firm initiated
Z-2117-2016Class IITerminatedVoluntary: Firm initiated
Z-2118-2016Class IITerminatedVoluntary: Firm initiated