Ingenuity Core 128-Computed Tomography X-ray system
Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to... on . FDA classified it as Class II and gives the reason as "Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2118-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes..
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- Philips Medical Systems (Cleveland) Inc. sent an Urgent Field Safety Notice letter to affected customers , via certified letter with receipts returned for tracking purposes. The letter identified the affected product, problem and actions to be taken. Philips Healthcare is implementing software update version 3.5.5 to correct the problem. For questions contact your local Philips representative or local Philips Healthcare office.
- Quantity in commerce
- 19
- Distribution
- US Nationwide Distribution in the states of CA, FL, ID, KS, MI, NY,OH,TX
- Product codes and lots
- Ingenuity Core128:System Code #728323, Serial numbers: 320001, 320016, 320076, 320077, 320080, 320085, 320086, 320087, 320088, 320089, 320091, 320113, 336001, 336002, 336003, 336006, 336007, 336008, 336009, 336010, 336014
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2116-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2117-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2118-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
