Target Detachable Coils: TARGET HELICAL NANO 2MM X 3CM MODEL Number:M0035452030 Neurology: Target Detachable...
Stryker Neurovascular began a recall of Target Detachable Coils: TARGET HELICAL NANO 2MM X 3CM MODEL Number:M0035452030 Neurology: Target Detachable Coils are intended to endovascularly obstruct or... on . FDA classified it as Class II and gives the reason as "Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2179-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Target Detachable Coils: TARGET HELICAL NANO 2MM X 3CM MODEL Number:M0035452030 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
- Product code
- HCG Device, Neurovascular Embolization
- Reason for recall
- Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- Stryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated June 3, 2016, to all affected customers via FED Ex. Letter includes recall letter, list of impacted lot numbers, customer acknowledgement form and frequently asked questions. Customers were instructed to immediately check their inventory. segregate the effected inventory in a secure location for return to Stryker, circulate the Field Safety Notice internally to all interested/affected parties. Inform Stryker if any of the subject devices have been distributed to other organizations. Customers were also instructed to complete the attached customer response form and return the completed form to their nominated Stryker Representative.
- Quantity in commerce
- 45,302 (8,188 in the U.S.) - total, all model numbers
- Distribution
- Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
- Product codes and lots
- MODEL Number::M0035452030; Lot Number/Expiration Date: 19032837 30-Apr-19 19034203 31-May-19 18953644 31-Mar-19 19032930 30-Apr-19 18986505 30-Apr-19 18922370 31-Mar-19 18987445 30-Apr-19 18972051 30-Apr-19 18971500 30-Apr-19 18905990 31-Mar-19 19060369 31-May-19.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2160-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2161-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2162-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2163-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2164-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2165-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2166-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2167-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2168-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2169-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2170-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2171-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2172-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2173-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2174-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2175-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2176-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2177-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2178-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2179-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2180-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2181-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2182-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2183-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2184-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2185-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2186-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2187-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.