510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2261-2016Class II

Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000...

The Metrix Company / initiated 2016-06-23 / Terminated / root cause: design

The Metrix Company began a recall of Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000 mL on . FDA classified it as Class II and gives the reason as "An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2261-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Metrix Company
Product
Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000 mL.
Product code
NEP System/Device, Pharmacy Compounding
Reason for recall
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
Root cause tag
design
FDA root cause
Device Design
Action
On 6/23/2016, the recalling firm sent written recall notices to their distributors and end users notifying them of the recall. Customers and consignees were instructed to not use the devices, locate and quarantine all affected devices, and to complete the response form.
Quantity in commerce
18,366 cases
Distribution
Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
Product codes and lots
All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 58740, 58741, 58742, 58749, 58750, 58753, 58754

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2261-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2257-2016Class IITerminatedVoluntary: Firm initiated
Z-2258-2016Class IITerminatedVoluntary: Firm initiated
Z-2259-2016Class IITerminatedVoluntary: Firm initiated
Z-2260-2016Class IITerminatedVoluntary: Firm initiated
Z-2261-2016Class IITerminatedVoluntary: Firm initiated
Z-2262-2016Class IITerminatedVoluntary: Firm initiated
Z-2263-2016Class IITerminatedVoluntary: Firm initiated
Z-2264-2016Class IITerminatedVoluntary: Firm initiated