510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2218-2016Class II

Smith and Nephew TWINFIX QUICK-T Knot Pusher Suture Cutter; Part Number: 7209417, for arthroscopic use

Smith & Nephew, Inc. / initiated 2016-06-15 / Terminated / root cause: packaging

Smith & Nephew, Inc. began a recall of Smith and Nephew TWINFIX QUICK-T Knot Pusher Suture Cutter; Part Number: 7209417, for arthroscopic use on . FDA classified it as Class II and gives the reason as "The device could be subject to breaches of its sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2218-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith and Nephew TWINFIX QUICK-T Knot Pusher Suture Cutter; Part Number: 7209417, for arthroscopic use.
Product code
NBH Accessories, Arthroscopic
Reason for recall
The device could be subject to breaches of its sterile packaging.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Urgent Recall Notification Letters were sent via Federal Express on June 15, 2016, to the customers listed in the firm's distribution report. Customers were asked to inventory their devices and contact the firm for return authorizations.
Quantity in commerce
563
Distribution
Worldwide distribution. US nationwide, AT, AU, DE, FR, GB, JP, PT, and RU.
Product codes and lots
Batch numbers: 50412843, 50477278, 50527184

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2218-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2218-2016Class IITerminatedVoluntary: Firm initiated