PoleStar system (N20 / N30) Warning Lamps Control Box, Warning Lamps Control Box Cables, Thermoflex Chiller...
Medtronic Navigation began a recall of PoleStar system (N20 / N30) Warning Lamps Control Box, Warning Lamps Control Box Cables, Thermoflex Chiller Product The PoleStar Surgical MRI System is... on . FDA classified it as Class II and gives the reason as "Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2129-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- PoleStar system (N20 / N30) Warning Lamps Control Box, Warning Lamps Control Box Cables, Thermoflex Chiller Product The PoleStar Surgical MRI System is intended for use as an intraoperative imaging device to produce MRI images of sections of the head selected by the physician.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user.
- FDA root cause
- Process change control
- Action
- Medtronic sent a Field Safety Notice letter dated August 1, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Medtronic will perform an inspection on site to verify and correct the grounding connection of the PoleStar System. For questions contact Medtronic Navigation Technical Services at 1-800-595-9709.
- Quantity in commerce
- 52
- Distribution
- Worldwide Distribution - US Nationwide.
- Product codes and lots
- Model No. 506A0107-01 & 506A0107-02 - Warning Lamps Control Box; 504A0276-01, 504A0277-01& 504A0278-01 - Warning Lamps Control Box Cables; 506A0297-01 & 506A0298-01 - Thermoflex Chiller
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2129-2016 | Class II | Terminated | Voluntary: Firm initiated |