Z-2129-2016Class II

PoleStar system (N20 / N30) Warning Lamps Control Box, Warning Lamps Control Box Cables, Thermoflex Chiller...

Medtronic Navigation, Inc / initiated 2016-06-20 / Terminated

Medtronic Navigation began a recall of PoleStar system (N20 / N30) Warning Lamps Control Box, Warning Lamps Control Box Cables, Thermoflex Chiller Product The PoleStar Surgical MRI System is... on . FDA classified it as Class II and gives the reason as "Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2129-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
PoleStar system (N20 / N30) Warning Lamps Control Box, Warning Lamps Control Box Cables, Thermoflex Chiller Product The PoleStar Surgical MRI System is intended for use as an intraoperative imaging device to produce MRI images of sections of the head selected by the physician.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user.
FDA root cause
Process change control
Action
Medtronic sent a Field Safety Notice letter dated August 1, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Medtronic will perform an inspection on site to verify and correct the grounding connection of the PoleStar System. For questions contact Medtronic Navigation Technical Services at 1-800-595-9709.
Quantity in commerce
52
Distribution
Worldwide Distribution - US Nationwide.
Product codes and lots
Model No. 506A0107-01 & 506A0107-02 - Warning Lamps Control Box; 504A0276-01, 504A0277-01& 504A0278-01 - Warning Lamps Control Box Cables; 506A0297-01 & 506A0298-01 - Thermoflex Chiller

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2129-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2129-2016Class IITerminatedVoluntary: Firm initiated