Z-2212-2016Class I

DIAMOND-FLEX CIRCULAR RETRACTOR, ANGLED, 40MM 5MM, Product Code 89-6114, Lot Code(s): 878971; 879366; 879404...

Medtronic, Inc. / initiated 2016-07-01 / Terminated / root cause: manufacturing process

Carefusion 2200 Inc began a recall of DIAMOND-FLEX CIRCULAR RETRACTOR, ANGLED, 40MM 5MM, Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16. Designed to retract or... on . FDA classified it as Class I and gives the reason as "BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were to occur while in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2212-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
DIAMOND-FLEX CIRCULAR RETRACTOR, ANGLED, 40MM 5MM, Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16. Designed to retract or elevate organs and tissue to provide better visualization access.
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were to occur while in use in a procedure it has the potential to damages tissue or organs
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
BD, sent an "URGENT: Medical Device/ Safety Alert/Recall Notification" letter dated 6/30/2016 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to return the affected units, along with the enclosed CUSTOMER RESPONSE FORM to: BD formerly CareFusion, 75 North Fairway Drive, Vernon Hills, Illinois, 60061 Attn: Customer Advocacy; to expedite the correction process and acknowledge receipt of the notification. The firm will issue a credit upon receipt of the returned affected product. For questions and support 1-800-323-9088 Prompt 3 or email: [email protected].
Quantity in commerce
10
Distribution
Worldwide Distribution - US, including the states of WA and MD; and, the country of Japan.
Product codes and lots
Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2212-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2212-2016Class ITerminatedVoluntary: Firm initiated