Z-2215-2016Class I
Oxylog 3000, Disposable Pediatric Patient Circuit, Catalog Number: 5704964 The VentStar Oxylog 3000F...
Draeger, Inc. / initiated 2016-05-31 / Terminated / root cause: design
Draeger Medical, Inc. began a recall of Oxylog 3000, Disposable Pediatric Patient Circuit, Catalog Number: 5704964 The VentStar Oxylog 3000F disposable pediatric patient breathing circuit is used... on . FDA classified it as Class I and gives the reason as "Isolated episodes of leakage at the check valve were found". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2215-2016
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Oxylog 3000, Disposable Pediatric Patient Circuit, Catalog Number: 5704964 The VentStar Oxylog 3000F disposable pediatric patient breathing circuit is used with the Drger Oxylog 3000 and Oxylog 3000 plus Emergency Transport Ventilators.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Isolated episodes of leakage at the check valve were found
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Drager sent an Urgent Medical Device Recall letter dated May 31, 2016 and Customer Reply and Order Card to all customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to inspect stock and dispose of affected products. For questions regarding this letter call 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of your Drger device/accessory contact DrgerService Technical Support at 1-800- 543-5047 (press 4 at the prompt).
- Quantity in commerce
- 42,922 units worldwide (US distribution 1,530)
- Distribution
- US Nationwide Distribution including states of: AK, CA, FL, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NY,OH,OK,OR,PA,TN,TX,UT,VA,WA, and WI.
- Product codes and lots
- Catalog Number: 5704964
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2215-2016 | Class I | Terminated | Voluntary: Firm initiated |