Z-2242-2016Class II
MODIFIED LINEAR STANDARD/LATERAL NECK TRIALS WITH SPRING SIDES REF S-200635 and S-200636
Encore Medical, LP / initiated 2016-06-21 / Terminated / root cause: manufacturing process
Encore Medical, Lp began a recall of MODIFIED LINEAR STANDARD/LATERAL NECK TRIALS WITH SPRING SIDES REF S-200635 and S-200636 on . FDA classified it as Class II and gives the reason as "Reported failure of the instrument. The retaining ring component of a neck trial became disassembled from the body of the neck". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2242-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- MODIFIED LINEAR STANDARD/LATERAL NECK TRIALS WITH SPRING SIDES REF S-200635 and S-200636
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Reported failure of the instrument. The retaining ring component of a neck trial became disassembled from the body of the neck.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- The recalling firm notified affected consignees via letter on 6/21/16. The consignees were instructed to return the affected devices.
- Quantity in commerce
- 14 (7 sets)
- Distribution
- CO, ID, TX
- Product codes and lots
- Lot: 313R1000, 313R1900, 313R1901, 314R1000, 314R1900, 314R1901
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2242-2016 | Class II | Terminated | Voluntary: Firm initiated |