Femoral Arterial Line Catheterization Kit with Sharps Safety Features 510K 810675, product code DQX, Device...
Arrow International Inc began a recall of Femoral Arterial Line Catheterization Kit with Sharps Safety Features 510K 810675, product code DQX, Device Listing D025180 Material ASK-04018-VUH Product... on . FDA classified it as Class II and gives the reason as "Labeling error. Some shipping cartons labeled as ASK-04018-VUH contained ASK-04020-VUH kits; and some shipping cartons labeled as ASK-04020-VUH contained ASK-04018-VUH kits". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2201-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Femoral Arterial Line Catheterization Kit with Sharps Safety Features 510K 810675, product code DQX, Device Listing D025180 Material ASK-04018-VUH Product Usage: The Arrow arterial catheterization device permits access to the peripheral arterial circulation.
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Labeling error. Some shipping cartons labeled as ASK-04018-VUH contained ASK-04020-VUH kits; and some shipping cartons labeled as ASK-04020-VUH contained ASK-04018-VUH kits.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Arrow International mailed an Urgent Medical Device Recall Notification letter and acknowledgement form dated June 20, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to 1. Immediately discontinue distribution and quarantine any products with the product codes and lot numbers listed above. 2. Inspect affected product within your control to identify if the shipping corrugate label correctly identifies the products inside the shipping carton. Products with an incorrect label on the shipping carton can be returned to Arrow. 3. Once you have completed inspection of all of the products from your own inventory, check the box on the enclosed Acknowledgment Form that indicates that you have completed the inspection and fax it to 1-855-419-8507, Attn: Customer Service, or email it to [email protected]. For questions contact your local sales representative or Customer Service at 1-866-246-6990.
- Quantity in commerce
- 50 units in total
- Distribution
- US distribution in the states of TN and KY.
- Product codes and lots
- Lot/Batch Numbers: 23F14M0829, 23F14M0986
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2201-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2202-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.