Z-2147-2016Class II

AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and...

LivaNova USA, Inc. / initiated 2016-06-22 / Terminated / root cause: manufacturing process

Sorin Group USA, Inc. began a recall of AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae are designed to be... on . FDA classified it as Class II and gives the reason as "Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2147-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm, LivaNova, sent an "Urgent Safety Alert" field communication letter dated June 22, 2016 via certified mail to consignees. The letter described the product, problem and actions to be taken. The consignees were instructed to: check your inventory to determine if you own Cannulae, if no inventory, complete the response form and return it per instructions indicated on form.; check the tip angle to determine if it is conforming to specifications, if specifications not met, file a complaint through your normal process whether you decide to keep it, use it or return it to Sorin Group; If you are not willing to use it, return affected product parts to Sorin Group USA,; contact Customer Service at 1-800-650-2623 or email CustomerService@livanova,com to have a RMA issued, and complete and return the Customer Response Form no later than July 20, 2016 via Fax at 303-467-6502 or by e-mail to [email protected]. If you need support for the execution of these actions, please contact your LivaNova sales representative or call Customer Service at 1-800-650-2623 (Monday to Friday 8AM to 5PM MT). For questions regarding this notice, please contact Customer Service at 1-800-650-2623 (Monday to Friday 8AM to 5PM MT) or e-mail [email protected].
Quantity in commerce
5740
Distribution
Worldwide Distribution: US (nationwide) to states of: LA, Ml, MO, OH, TN, and TX; and countries of: Italy and Canada.
Product codes and lots
Lots: 1505600017 ; 1505600018 ; 1506500005 ; 1509200078 ; 1513100112 ; 1513900027 ; 1513900028 ; 1514700045 ; 1516600192 ; 1516800126 ; 1517800004 ; 1519500037 ; 1520100049 ; 1520100050 ; 1520900031 ; 1520900033 ; 1521600126 ; 1523000093 ; 1523600120 ; 1525100023 ; 1525800026 ; 1526500032 ; 1526500033 ; 1526500034 ; 1526500072 ; 1527200050 ; 1527200051 ; 1527200052 ; 1527800163 ; 1529300026 ; 1529300027 ; 1529300028 ; 1533500033 ; 1534200034 ; 1601900027 ; 1601900028 ;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2147-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2147-2016Class IITerminatedVoluntary: Firm initiated