Z-2281-2016Class II

Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile...

Howmedica Osteonics Corp / initiated 2016-06-03 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032... on . FDA classified it as Class II and gives the reason as "Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2281-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032. For use in total hip arthroplasty.
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Stryker notified its branches/agencies of this recall by e-mail on May 26, 2016. They were asked to quarantine the affected devices. Stryker sent a recall letter and business reply form dated June 3, 2016 to their affected customers via UPS. Stericycle will be handling the returns.
Quantity in commerce
1,864 units
Distribution
Worldwide distribution. US nationwide (AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OH, PA, TX, WA, WI, WV), Sweden, South Africa, Chile, Australia, New Zealand, Canada, United Kingdom, Switzerland, Colombia, and Poland.
Product codes and lots
Catalog Nos. 2199-2022, 2199-2028 and 2199-2032 All lots manufactured from May 30, 2012

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2281-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2281-2016Class IITerminatedVoluntary: Firm initiated