Proadvantage Mono Rapid Test Cassette, Catalog No. P080016 a simple test that utilizes an extract of bovine...
Alere San Diego, Inc. began a recall of Proadvantage Mono Rapid Test Cassette, Catalog No. P080016 a simple test that utilizes an extract of bovine erythrocytes to qualitatively and selectively... on . FDA classified it as Class III and gives the reason as "The Positive Control information referencing human plasma as the antibody material described in the product labeling and packaging is incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2269-2016
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Proadvantage Mono Rapid Test Cassette, Catalog No. P080016 a simple test that utilizes an extract of bovine erythrocytes to qualitatively and selectively detect IM heterophile antibodies in whole blood, serum or plasma in just minutes.
- Product code
- KTN System, Test, Infectious Mononucleosis
- Reason for recall
- The Positive Control information referencing human plasma as the antibody material described in the product labeling and packaging is incorrect.
- Root cause tag
- labeling and UDI
- FDA root cause
- Unknown/Undetermined by firm
- Action
- A Medical Device Correction letter dated 6/22/16 will be sent to all customers to inform them that Alere is recalling the MONO Rapid Test products because the Positive Control Information referencing human plasma as the antibody material described in the product labeling and packaging is incorrect. Customers are informed that the labeling states the Positive Control contains the following: Diluted human plasma containing IM heterophile antibodies, 0.09% sodium azide. The Positive Control contains the following: Goat anti-mono antibody, 0.09% sodium azide. Customers are instructed to complete and fax or email the enclosed verification form within 10 days. Customers with questions are instructed to contact Alere San Diego at (844) 804-4945 or email at [email protected].
- Distribution
- US
- Product codes and lots
- all lots are affected
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2265-2016 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-2266-2016 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-2267-2016 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-2268-2016 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-2269-2016 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-2270-2016 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-2271-2016 | Class III | Terminated | Voluntary: Firm initiated |