Z-2879-2016Class II

Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the...

CareFusion 303, Inc / initiated 2016-08-08 / Terminated / root cause: software

CareFusion 303, Inc. began a recall of Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers fluids in a... on . FDA classified it as Class II and gives the reason as "A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2879-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers fluids in a manner similar to current syringe pumps on the market. Up to four Alaris Syringe pump modules can be connected to the Alaris PC unit which is the central programming, monitoring and power supply component for the Alaris System. The syringe pump uses standard, single-use administration sets and syringes with luer-lock connectors.
Product code
FRN Pump, Infusion
Reason for recall
A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another.
Root cause tag
software
FDA root cause
Software design
Action
The firm, BD, sent an "Urgent: Medical Device Recall Notification" letter dated 8/8/16 to customers to inform them that BD has identified a software anomaly associated with the Alaris Syringe module software version 9.15. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the instructions and promptly complete and return the Customer Response Card to expedite the corrective action process. Customers with recall related questions are instructed to contact the BD Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact customer advocacy at (888) 812-3266. Customers with technical questions regarding the Alaris System are instructed to contact Technical Support at (888) 812-3229.
Quantity in commerce
12,000 units
Distribution
Worldwide Distribution: US(nationwide) and countries of: Canada, Australia and United Arab Emirates.
Product codes and lots
Software version 9.15

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2879-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2879-2016Class IITerminatedVoluntary: Firm initiated