Z-2159-2016Class II

Merge OfficePACS software. The firm name on the label is Merge Healthcare, Heartland, WI

Merge Healthcare Incorporated / initiated 2016-01-31 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge OfficePACS software. The firm name on the label is Merge Healthcare, Heartland, WI on . FDA classified it as Class II and gives the reason as "Potential data loss occurs as a result of product archiving not working properly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2159-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge OfficePACS software. The firm name on the label is Merge Healthcare, Heartland, WI.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Potential data loss occurs as a result of product archiving not working properly.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued a letter dated 1/27/2016 via e-mail on 1/31/2016 explaining the reason for recall and instructing the customer to provide a copy of the notification to all users.
Quantity in commerce
49 sites have the affected software
Distribution
Distribution was made to medical facilities located in AL, AZ, CA, CO, CT, FL GA, ID, IL, KY, MA, MO, NE, NH, NJ, NY, NC, OK, PA, SC, TN, VA, and WA. There was no government/military/foreign distribution.
Product codes and lots
Version 3.4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2159-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2159-2016Class IITerminatedVoluntary: Firm initiated