Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera...
Karl Storz Endoscopy began a recall of Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1... on . FDA classified it as Class II and gives the reason as "Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2134-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Karl Storz Endoscopy-America, Inc.
- Product
- Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Video Mediastinoscope
- Product code
- FET Endoscopic Video Imaging System/Component, Gastroenterology-Urology
- Reason for recall
- Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- KARL STORZ sent an Customer Notification letter dated June 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. KARL STORZ has become aware become aware of a potential electrical interference issue when the lmage1 SPIES X-LINK and lmage1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinoscope and an electro-surgery generator (ESG). Customers are instructed to complete the attached acknowledgement Form and to return it to KARL STORZ. Customers with questions are instructed to contact Technical Support Dept. at (800) 421-0837, press 2, and then press 1.
- Quantity in commerce
- 619 units total (90 units in US)
- Distribution
- Worldwide Distribution - US Distribution to the states of : IL, NM, TX, MD, TN, PA, FL, KY, MO, NY, MN, DC, MA, WI, CT, MI, ID, CO, GA, NJ and AR., and to the countries of : Argentina, Australia, Belgium, Bulgaria, China, Germany, Finland, Great Britain, India, Italy and Japan.
- Product codes and lots
- NT743473-P NT743472-P NT742840-P XU738911-P YU741045-P XU738248-P WU737260-P PV715203-P PV715207-P NT742829-P XU738899-P WU737275-P YV724235-P WU737269-P WU737268-P VU736411-P XU738901-P XU738910-P ZU742677-P YU741050-P VU736405-P TU733422-P TU733419-P SU733063-P VU736390-P VU736386-P UU734392-P RV716122-P OT744400-P NT743287-P OT744395-P NU727551-P NU727552-P NU727706-P TU734218-P TU734217-P NT742846-P YU741046-P OU728667-P TU733735-P VU736388-P NT742848-P NT742838-P OU728672-P OU728427-P WU737813-P OU729135-P OU728431-P OU728430-P WU737805-P ZU742670-P ZU742671-P YU741049-P YU740475-P ZU742681-P RV716116-P SU733055-P SU733057-P SU733059-P SU733060-P TU733418-P TU733420-P TU733421-P TU733423-P TU733424-P TU733425-P TU733427-P WU737810-P XU738914-P YU740004-P YU740471-P YU740479-P YU740480-P YU741036-P NV711744-P PV715216-P OU728670-P NU726430-P OT744399-P OT744389-P OT744386-P YU739999-P YU740003-P YU740005-P UU734395-P YV724443-P YV724442-P YV724440-P YV724438-P YV724245-P
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2134-2016 | Class II | Terminated | Voluntary: Firm initiated |