Z-2295-2016Class II

Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010)...

Draeger, Inc. / initiated 2016-06-22 / Terminated / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software... on . FDA classified it as Class II and gives the reason as "Risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2295-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled. Product Usage: Babytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units. Therapy and nursing uses: - As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
Product code
FMT Warmer, Infant Radiant
Reason for recall
Risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Drager mailed to customers an Urgent Medical Device Recall Notification Letter dated June 22, 2016 to affected customers. The letter identified the affected product, problem information on how to operate the device and replacement of parts information. For questions regarding this letter please call 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of your Drger Babytherm 8004/8010 please contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt).
Quantity in commerce
1,599 units
Distribution
US Nationwide Distribution including Puerto Rico and the states of AK,AL,AR,CA,CO,CT,DE,FL,GA,HI,IA,IL,IN,KS,KY,LA,MA,MD,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,TN,TS,TX,UT,VA,WA,WI,WV, and WY.
Product codes and lots
Serial Numbers: ARLN-xxxx to ASFE-xxxx Catalog Numbers: FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2295-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2295-2016Class IITerminatedVoluntary: Firm initiated