Z-2299-2016Class II

ADVIA Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent is for in vitro diagnostic use in the...

Siemens Healthcare Diagnostics Inc. / initiated 2016-06-17 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent is for in vitro diagnostic use in the quantitative analysis of theophylline in human serum or... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has confirmed the Syva EMIT 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2299-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent is for in vitro diagnostic use in the quantitative analysis of theophylline in human serum or plasma on the ADVIA Chemistry Systems. Copies of IFUs included in submission.
Product code
KLS Enzyme Immunoassay, Theophylline
Reason for recall
Siemens Healthcare Diagnostics has confirmed the Syva EMIT 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for commercially available Bio-Rad Liquichek TDM Quality Control Levels and patient samples
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens Healthcare Diagnostics mailed a letter to customers on June 17, 2016, asking that they discontinue use of and discard the affected kit lots identified in the recall notification letter. Customers were also asked to complete and return the Field Correction Effectiveness Check/Product Replacement Form that was mailed with the Recall Notification Letter.
Quantity in commerce
658 units
Distribution
Distributed to: CA, CT, DE, FL, GA, IA, IN, KS, KY, LA, MA, MD, MI, MO, NC, NJ, NM, NY, OH, OK, PA, TN, VA, WA
Product codes and lots
ADVIA¿ Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent - SMN# 10377503, Lot # 334018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2299-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2298-2016Class IITerminatedVoluntary: Firm initiated
Z-2299-2016Class IITerminatedVoluntary: Firm initiated