Z-2252-2016Class II

BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absolute...

Becton, Dickinson and Company, BD Biosciences / initiated 2016-06-24 / Terminated / root cause: packaging

Becton, Dickinson and Company, BD Biosciences began a recall of BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absolute counts of leucocytes in blood on . FDA classified it as Class II and gives the reason as "One lot of BD Trucount Absolute Counting Tubes (IVD) may not be sealed properly and exposure to ambient environment can rapidly degrade the performance of the product". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2252-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absolute counts of leucocytes in blood.
Product code
GKZ Counter, Differential Cell
Reason for recall
One lot of BD Trucount Absolute Counting Tubes (IVD) may not be sealed properly and exposure to ambient environment can rapidly degrade the performance of the product.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Urgent Product Correction notifications letters were sent on June 24, 2016 to all affected customers. Letters identify the issue and product and provide illustrations of the Kit Box and pouch labels. Correction letters identified the issue and the product involved with photos of product labels. BD letters states that they want to reinforce the need to visually inspect the desiccant contained in each pouch . If the desiccant has turned from blue to lavender, discard the remaining tubes. The acknowledgement form should be completed and returned regardless if any inventory remains. If you require further assistance, please contact BD Customer Support at 855.236.2772 (prompt 3) in the Unites States. For customers outside the US, contact your local BD Biosciences representative or distributor
Quantity in commerce
328 US, 516 International
Distribution
USA Distribution to the states of : NC, WA, WI, NJ, PR, MD, NY, MA, CA, NY , MD, KY, IN, VA, SD, PA, OH, AR, AZ and Foreign to : Canada, Belgium and Chile
Product codes and lots
Catalog 340334, Lot ;6083708 (embossing ID of (160216T3)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2252-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2252-2016Class IITerminatedVoluntary: Firm initiated