Z-0138-2017Class II
OER-Pro Endoscope Reprocessor, Endoscope washer/disinfector
OLYMPUS Corporation of the Americas / initiated 2016-06-23 / Terminated
Olympus Corporation of the Americas began a recall of OER-Pro Endoscope Reprocessor, Endoscope washer/disinfector on . FDA classified it as Class II and gives the reason as "New warning requiring Users to conduct duodenoscope precleaning and manual cleaning even when using an Automated Endoscope Reprocessor (AER) that has instructions that may indicate a user could forego certain steps in...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0138-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- OER-Pro Endoscope Reprocessor, Endoscope washer/disinfector
- Product code
- FEB Accessories, Cleaning, For Endoscope
- Reason for recall
- New warning requiring Users to conduct duodenoscope precleaning and manual cleaning even when using an Automated Endoscope Reprocessor (AER) that has instructions that may indicate a user could forego certain steps in precleaning and manual cleaning of the endoscopes.
- FDA root cause
- Under Investigation by firm
- Action
- Olympus America Inc. (OAI) mailed a letter to customers informing them that they are issuing an updated Operation Manual for the OER-Pro Automated Endoscope Reprocessor. The following actions were asked to be taken: -Olympus has discontinued previously distributed copies of the OER-Pro Operation Manual. - Inspect your inventory of Operation Manuals and discard any existing inventory of OER-Pro Operation Manuals. - Implement use of the enclosed OER-Pro Operation Manual, which contains the new Warning on pages 73 and 80. - Ensure all reprocessing personnel are completely knowledgeable and thoroughly trained on the requirement to perform precleaning and manual cleaning of Olympus duodenoscopes prior to placing the duodenoscopes in the OER-Pro for high level disinfection. - Additional copies of the new OER-Pro Operation Manual can be obtained by contacting our Technical Assistance Center at 1-800-848-9024, option 1, or by indicating on the enclosed questionnaire. Additional Operation Manuals will be mailed to your facility - Please indicate on the enclosed questionnaire that you have received this notification. Fax the completed form to (484) 896-7128.
- Quantity in commerce
- 2, 686 units
- Distribution
- Distributed to: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
- Product codes and lots
- Model: OER-Pro, All serial numbers affected
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0138-2017 | Class II | Terminated | Voluntary: Firm initiated |