Z-2343-2016Class II

91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 &...

GE Healthcare / initiated 2016-07-01 / Terminated / root cause: supplier

Spacelabs Healthcare Inc began a recall of 91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration... on . FDA classified it as Class II and gives the reason as "The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-2343-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP. Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP. Option I: SpO2, Temperature x2, & adult/neonate NIBP.
Product code
DSI Detector And Alarm, Arrhythmia
Reason for recall
The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
Spacelabs expanded the recall and sent an Urgent - Medical Device Correction letter dated August 30, 2016, to additional affected customers via priority mail (return receipt requested). On September 6, 2016, a customer letter (translated as necessary) was sent via email to all international subsidiaries and distributors of record. ********************************************************************************************** Spacelabs Healthcare Inc, sent an Urgent - Medical Device Correction letter dated July 1, 2016, to all affected customers (return receipt requested). On July 8, 2016, a customer letter (translated as necessary) was sent via email to all international subsidiaries and distributors of record. Spacelabs Healthcare will contact customers to schedule a convenient time for a Spacelabs service technician to update their affected Command Module(s) at no cost. For additional information or technical assistance, please contact: Technical Support 1-800-522-7025 and select (2) for Technical Support
Quantity in commerce
3801 modules and 116 PCBAs & valves distributed in the US; 1297 modules and 366 PCBAs & valves distributed outside the US
Distribution
Worldwide Distribution - USA Distribution and to the countries of : AFGHANISTAN, ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BRUNEI DARUSSALAM; CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DOMINICAN REPUBLIC; ECUADOR, EGYPT, ESTONIA, FRANCE, GERMANY, GUATEMALA, HONG KONG, INDIA, IRAN (ISLAMIC REPUBLIC OF), ISRAEL, ITALY, JORDAN, KUWAIT, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, PAKISTAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, Slovakia, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VIET NAM.
Product codes and lots
Manifold Lot Codes: 128280, 129420, 129940, or 155640.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2343-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2343-2016Class IITerminatedVoluntary: Firm initiated