Z-2645-2016Class II

DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket...

Medtronic Neuromodulation / initiated 2016-06-23 / Terminated / root cause: manufacturing process

Medtronic Neuromodulation began a recall of DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can be used as a part of a neurostimulation... on . FDA classified it as Class II and gives the reason as "Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2645-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can be used as a part of a neurostimulation system for deep brain stimulation. The pocket adaptor is intended to be implanted with the new replacement neurostimulator in the same pocket used for the explanted neurostimulator. Implanting in the same neurostimulator pocket allows for a single-incision procedure.
Product code
MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
Reason for recall
Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were sent on 6/23/2016 a "Medical Device Safety Notification," dated June 2016. The letter was addressed to Healthcare Professional. The letter described "Background" information on the problem and product involved in the recall. It also provided "Recommendations" which included "Implanting the pocket adaptor with the neurostimulator" and "Checking system integrity". For questions contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 weekdays 7 am - 6pm CST.
Quantity in commerce
7,542 in the US and 12,999 outside the US.
Distribution
Worldwide Distribution.
Product codes and lots
Notification is not lot specific.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2645-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2645-2016Class IITerminatedVoluntary: Firm initiated