510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2391-2016Class II

Coolife Sinergy Radiofrequency Kit, SIK-17-75-4 (109971504); Kit used in conjunction with a Radiofrequency...

Halyard Health / initiated 2016-06-16 / Terminated / root cause: manufacturing process

Halyard Health, Inc began a recall of Coolife Sinergy Radiofrequency Kit, SIK-17-75-4 (109971504); Kit used in conjunction with a Radiofrequency Generator to create radiofrequency lesions in... on . FDA classified it as Class II and gives the reason as "Mismatch between the length of the RF electrode (probe) and the cannula (introducer)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2391-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
Coolife Sinergy Radiofrequency Kit, SIK-17-75-4 (109971504); Kit used in conjunction with a Radiofrequency Generator to create radiofrequency lesions in nervous tissue.
Product code
GXI Probe, Radiofrequency Lesion
Reason for recall
Mismatch between the length of the RF electrode (probe) and the cannula (introducer),
Root cause tag
manufacturing process
FDA root cause
Process control
Action
All distributors and healthcare facilities that were directly shipped recalled devices were notified of the recall beginning 6/16/2016 by email and by FedEx. They were instructed to examine their stock and to quarantine and discontinue use of the product. The letter contained a response form which was to faxed back to Halyard.
Quantity in commerce
51 kits
Distribution
Distributed in the states of CT, IL, MA, MI, NV, NH, NC, and TX, and the country of Canada.
Product codes and lots
Lot Number M5264D205

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2391-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2391-2016Class IITerminatedVoluntary: Firm initiated