Ingenuity Core, model # 728326; Intended to produce cross- sectional images of the body
Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity Core, model # 728326; Intended to produce cross- sectional images of the body on . FDA classified it as Class II and gives the reason as "Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2115-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Ingenuity Core, model # 728326; Intended to produce cross- sectional images of the body.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Philips is informing the consignee of this issue through: Customer Information Letter (CIL) 72800656 . International consignees will be contacted by Philips Healthcare Global Market representatives, who will distribute the CIL. A Field Service Engineer (FSE) will contact the customers to schedule a time to conduct the appropriate correction through Field Change Order (FCO) 72800656 once it is available. This FCO will address the above identified issue. A response form sent to the customer via certified mail. This package will also contain a return receipt.
- Quantity in commerce
- 1 unit
- Distribution
- Distributed in China, Denmark, Germany, Hungary, Italy, Philippines, Spain, Sweden, Switzerland, Thailand, and United Kingdom.
- Product codes and lots
- 901510
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2111-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2112-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2113-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2114-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2115-2016 | Class II | Terminated | Voluntary: Firm initiated |