AU5800 Chemistry Analyzers, Reference No. A98352 The AU series are automated, clinical chemistry analyzers...

Siemens Healthcare Diagnostics Inc. / initiated 2015-07-30 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of AU5800 Chemistry Analyzers, Reference No. A98352 The AU series are automated, clinical chemistry analyzers that measure analytes in patient samples (serum... on . FDA classified it as Class II and gives the reason as "Beckman coulter has received complaints regarding leaking and cracked C cups. When the cup cracks, coolant can leak onto the connectors of the Printed Circuit Boards (PCB) in the analyzer, resulting in a short-circuit...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2310-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AU5800 Chemistry Analyzers, Reference No. A98352 The AU series are automated, clinical chemistry analyzers that measure analytes in patient samples (serum, plasma, urine, CSF) in combination with appropriate reagents, calibrators, quality, control (QC) materials, and other accessories.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman coulter has received complaints regarding leaking and cracked C cups. When the cup cracks, coolant can leak onto the connectors of the Printed Circuit Boards (PCB) in the analyzer, resulting in a short-circuit causing the instrument to go into a stop mode.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
In March 2011, the material and molding process of the Coolant cup (C cup) was changed. BEC did not issue a customer notification letter, as we do not believe that a notificationletter was either warranted or of any potential use or value to our customers, because the coolant cup (C cup) is an internal component that only a BEC Field Service Engineer (FSE) can access and inspect i.e., the C cup is not in a customer-accessible part of the analyzer. Thus, there are no specific actions the customer could take to inspect, identify, or mitigate the potential for a crack or leakage of the C cup. For questions regarding this recall call 714-961-4933.
Quantity in commerce
2,130 units total (588 units in US)
Distribution
Worldwide Distribution - US (nationwide (including Guam and Puerto Rico), and Internationally to Albania, Algeria, Australia, Austria, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Finland, France, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Lebanon, Liberia, Lithuania, Malawi, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States Uruguay, Viet Nam, Zambia, Zimbabwe.
Product codes and lots
1040016

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2310-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2308-2016Class IITerminatedVoluntary: Firm initiated
Z-2309-2016Class IITerminatedVoluntary: Firm initiated
Z-2310-2016Class IITerminatedVoluntary: Firm initiated