UniCel DxC Synchron Clinical Systems and UniCel DxC Synchron Access Integrated Clinical Systems The affected...

Siemens Healthcare Diagnostics Inc. / initiated 2014-04-22 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of UniCel DxC Synchron Clinical Systems and UniCel DxC Synchron Access Integrated Clinical Systems The affected part is Part number 70300. Part number 970300 is... on . FDA classified it as Class II and gives the reason as "Possible leaking of No Foam solution from the No Foam Bottle Assembly on the UniCel DxC Synchron Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2272-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
UniCel DxC Synchron Clinical Systems and UniCel DxC Synchron Access Integrated Clinical Systems The affected part is Part number 70300. Part number 970300 is used to transfer No Foam Solution from the No Foam Bottle to the Waste B canister and the Waste Manifold on the DxC Hydro System. Catalog No. UniCel DxC 800 Synchron Clinical Systems UniCel DxC 600 Synchron Clinical Systems UniCel DxC 600i Synchron Access Integrated Clinical Systems UniCel DxC 880i Synchron Access Integrated Clinical Systems UniCel DxC 660i Synchron Access Integrated Clinical Systems Product Labeling Reference No. A93719 For the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Possible leaking of No Foam solution from the No Foam Bottle Assembly on the UniCel DxC Synchron Systems.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Important Product Notice dated 4/22/14 was sent to all their customers to inform them that Beckman Coulter has identified a manufacturing issue with a plastic fitting on the No Foam Bottle Assembly which may crack and cause No Foam solution to leak. The letter informs the customers of the impact, action to be taken, and resolution. Customers are asked to inspect the No Foam Bottle Assembly for leaks prior to refilling the bottle. If leaks occur, customers should contact Customer Support Center following the instructions provided. Customers are instructed to complete the enclosed response form within 10 days. Customers with questions are instructed to contact the firm's Customer Support Center - Via our website, http://www.beckmancoulter.com/customersupport/support - Via phone, call 1-800-854-3633 in the United States and Canada - Outside the United States and Canada, contact your local Beckman Coulter Representative.
Quantity in commerce
3,540 units total (1,067 units in US)
Distribution
Worldwide Distribution -- United Arab Emirates, Angola, Australia, Belgium, Bulgaria, Canada, Switzerland, China, Colombia, Costa Rica , Czech Republic, Germany, Denmark, Egypt, Eritrea, Spain, Finland, France, United Kingdom, Greece, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Jordan, Japan, Korea, Republic of, Kuwait, Lebanon, Lithuania, Morocco, Mexico, Malaysia, Namibia, Netherlands, Norway, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Romania, Serbia, Russian Federation, Rwanda, Saudi Arabia, Sweden, Singapore, Slovakia, Swaziland, Thailand, Tunisia, Turkey, Taiwan, US (including Puerto Rico), Vietnam, South Africa, and Zimbabwe.
Product codes and lots
Up to and including 6133

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2272-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2272-2016Class IITerminatedVoluntary: Firm initiated