Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone...
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone cement. Stryker AutoPlex System kits contain a... on . FDA classified it as Class II and gives the reason as "The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2536-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone cement. Stryker AutoPlex System kits contain a cement mixer, cement injector, funnel, and an extension tube assembly whereby bone cement powder and liquid monomer are mixed inside a chamber and transferred into a cement injector automatically. The bone cement is then dispensed directly into a fracture site using the cement injector and introducer needle combination.
- Product code
- NDN Cement, Bone, Vertebroplasty
- Reason for recall
- The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs to be prepared for the injection procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 6/13/2016, URGENT MEDICAL DEVICE RECALL NOTIFICATION notifications were sent to the affected customers via courier and sales representatives via email. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Jennifer Olson 269-389-2644 [email protected]
- Quantity in commerce
- 4,841 boxes (9,682 units)
- Distribution
- Worldwide distribution. US nationwide, France, Netherlands, Spain, Italy, Canada, Israel, Romania, Switzerland, and United Kingdom.
- Product codes and lots
- Model Number(s): 0607-687-000 0605-887-000 0605-687-000 0605-683-000 0607-687-000 Lot numbers: 16021012, 16027022, 16049012, 16061012, 16092012, 16107012, 16022012, 16028012, 16049022, 16063012, 16098012, 16109012, 16025012, 16028022, 16049032, 16063022, 16100012, 16109022, 16025022, 16033012, 16053012, 16070012, 16104012, 16113022, 16026012, 16041012, 16056012, 16070022, 16104022, 16113032, 16026022, 16048012, 16056022, 16074012, 16105012, 16118012, 16027012, 16048022, 16060012, 16088012, 16106012, 16118022 0605-887-000 Lot numbers: 16041012, 16040012, 16050012, 16057012, 16069012, 16078012, 16104012, 16112012, 16124012 0605-687-000 Lot numbers: 16041012, 16047012, 16111012 0605-683-000 Lot numbers: 16071012
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2536-2016 | Class II | Terminated | Voluntary: Firm initiated |