Integra External Fixation System Rocker Bottom Model # 12225400. For use in the treatment of bone conditions...
Integra LifeSciences Corp. began a recall of Integra External Fixation System Rocker Bottom Model # 12225400. For use in the treatment of bone conditions amenable to treatment by the use of external... on . FDA classified it as Class II and gives the reason as "The silicone insert embedded in the rocker bottom has the potential to become dislodged or loose from the rocker bottom mold". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2296-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Integra External Fixation System Rocker Bottom Model # 12225400. For use in the treatment of bone conditions amenable to treatment by the use of external fixation.
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- The silicone insert embedded in the rocker bottom has the potential to become dislodged or loose from the rocker bottom mold.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- All affected consignees were notified via an "Urgent: Voluntary Medical Device Recall" letter sent on 7/1/16. The letter identified the affected product as well as the reason for the recall. Customers were asked to check their traceability records for shipments of the affected product. If the affected parts are found, customers are to stop use and distribution. A copy of the letter should be provided to any customers that the product may have been distributed to. The attached "Acknowledgement and Return Form" should be completed and returned regardless if customers have affected product or not. Questions regarding the instructions should be directed to Customer Service at 1-800-654-2873.
- Quantity in commerce
- 27 units
- Distribution
- Distributed in the US to the states of FL, VA, OK, CO, DC, and IN.
- Product codes and lots
- Lot # PN625.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2296-2016 | Class II | Terminated | Voluntary: Firm initiated |