Z-2296-2016Class II

Integra External Fixation System Rocker Bottom Model # 12225400. For use in the treatment of bone conditions...

Integra LifeSciences Corporation / initiated 2016-07-01 / Terminated / root cause: design

Integra LifeSciences Corp. began a recall of Integra External Fixation System Rocker Bottom Model # 12225400. For use in the treatment of bone conditions amenable to treatment by the use of external... on . FDA classified it as Class II and gives the reason as "The silicone insert embedded in the rocker bottom has the potential to become dislodged or loose from the rocker bottom mold". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2296-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra External Fixation System Rocker Bottom Model # 12225400. For use in the treatment of bone conditions amenable to treatment by the use of external fixation.
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
The silicone insert embedded in the rocker bottom has the potential to become dislodged or loose from the rocker bottom mold.
Root cause tag
design
FDA root cause
Device Design
Action
All affected consignees were notified via an "Urgent: Voluntary Medical Device Recall" letter sent on 7/1/16. The letter identified the affected product as well as the reason for the recall. Customers were asked to check their traceability records for shipments of the affected product. If the affected parts are found, customers are to stop use and distribution. A copy of the letter should be provided to any customers that the product may have been distributed to. The attached "Acknowledgement and Return Form" should be completed and returned regardless if customers have affected product or not. Questions regarding the instructions should be directed to Customer Service at 1-800-654-2873.
Quantity in commerce
27 units
Distribution
Distributed in the US to the states of FL, VA, OK, CO, DC, and IN.
Product codes and lots
Lot # PN625.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2296-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2296-2016Class IITerminatedVoluntary: Firm initiated