Optical Surface Monitoring System{Align RT Plus} This is used on Varian's TrueBeam EDGE; Radiology: The...

Varian Medical Systems, Inc. / initiated 2016-06-23 / Terminated / root cause: software

Varian Medical Systems, Inc. began a recall of Optical Surface Monitoring System{Align RT Plus} This is used on Varian's TrueBeam EDGE; Radiology: The AlignRT Plus system is indicated for use to position... on . FDA classified it as Class II and gives the reason as "Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2150-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Optical Surface Monitoring System{Align RT Plus} This is used on Varian's TrueBeam EDGE; Radiology: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to Withhold the beam automatically during radiation delivery.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different couch rotation from the plan couch rotation while performing patient alignment with Optical Surface Monitoring System [OSMS].
Root cause tag
software
FDA root cause
Software design
Action
The firm, Visionrt, sent an "URGENT MEDICAL DEVICE CORRECTION NOTIFICATION-URGENT FIELD SAFETY NOTIFICATION" letter dated June 24, 2016 to all affected customers. Letters describe the product, problem and actions to be taken. The customers were instructed as follows: DO NOT USE the "Move Couch" function if the couch rotation value set in the AlignRT device DOES NOT match the planned couch rotation for the treatment field; retain a copy of this document along with your most current product labeling; pass the letter to all those who need to be aware of it within their organization; always confirm that the patient is correctly aligned prior to commencing treatment; is not required to return their devices; promptly inform Vision RT if they believe that patient harm occurred, and complete Appendix 1- Customer Acknowledgment Response and return it via email to [email protected]. Should you have any queries on this letter, please do not hesitate to contact Vision RT by telephone on +44 20 83464300 (866-778-2379 from the US) or as per http://www.visionrt.com/contact/details.
Quantity in commerce
21 AlignRT Plus in US, 1 International.
Distribution
Worldwide Distribution: US (nationwide) and countries of: Austria, Azerbaijan, Brunei Darussalam, Bulgaria, Denmark, Georgia, Germany, India, Italy, Malaysia, Morocco, Norway, Portugal, Switzerland, and United Kingdom.
Product codes and lots
TrueBeam version 2.0/2.5 with any version OSMS.  AlignRT Serial Numbers: 247-0200 247-0209 248-0013 248-0057 248-0070 249-0057 249-0066 249-0163 249-0164 249-0195 249-0226 249-0274 249-0050 249-0286 249-0277 248-0044 249-0227 247-0208 249-0047 247-0218 248-0050 248-0043 248-0048 249-0293 249-0155.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2150-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2149-2016Class IITerminatedVoluntary: Firm initiated
Z-2150-2016Class IITerminatedVoluntary: Firm initiated