510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2205-2016Class II

6.0mm Round Fluted Bur, Super Long

Stryker Corp. / initiated 2016-06-16 / Terminated

Stryker Instruments Div. of Stryker Corporation began a recall of 6.0mm Round Fluted Bur, Super Long on . FDA classified it as Class II and gives the reason as "Stryker Instruments initiated a voluntary recall of specific lots of Round Fluted Burs, due to tarnishing or corrosion which may be present on the recalled burs and could result in a foreign body reaction (inflammation)...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2205-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
6.0mm Round Fluted Bur, Super Long
Product code
HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Reason for recall
Stryker Instruments initiated a voluntary recall of specific lots of Round Fluted Burs, due to tarnishing or corrosion which may be present on the recalled burs and could result in a foreign body reaction (inflammation) necessitating surgical intervention.
FDA root cause
Under Investigation by firm
Action
Stryker initiated a voluntary recall of the Stryker Round Fluted Burs (used with Stryker CORE" System) on June 16, 2016, via certified mail due to the potential for the devices to have tarnishing or corrosion present. Non-responding accounts will be contacted by phone, email, and/or fax as necessary in an effort to obtain signed Business Reply Forms. Affected product will be removed from the field. Further corrective and preventative actions will be determined through the associated CAPA. The existing label will be defaced, packaging will be opened and products will be destroyed. Actions to be taken by the Customer/User: 1. Immediately review this Recall Notification. 2. Check all stock areas and/or operating room storage to determine how many affected Round Fluted Burs from the affected lots are at your facility. Quarantine and discontinue use of the recalled Round Fluted Burs. 3. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many, if any, affected items are currently in your inventory. Please complete and return the BRF even if you dont have any affected product on hand. Note: Your signature on the BRF indicates that you received and understand this Notification and have followed the instructions in the Notification. 4 If you have further distributed this product, please forward this letter and the attached BRF to all affected locations. Please indicate each location on the BRF. 5. Fax the completed BRF to Stryker Instruments Regulatory Department at 866-521-2762. 6. If the BRF for your facility indicates that recalled product is currently on hand, a FedEx label will be emailed to you. This shipping label should be used to return recalled product. Upon receipt of the recalled product, a credit will be issued to your account. Report any serious adverse events or product quality problems to Stryker Instruments: 1-800-253-3210 Health care professionals and consumers
Quantity in commerce
25
Distribution
US Distribution to the states of : CA, IL, SD Foreign: Canada, Netherlands
Product codes and lots
5190-010-060 Lot Numbers 15231017, 16033017, 16066017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2205-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2204-2016Class IITerminatedVoluntary: Firm initiated
Z-2205-2016Class IITerminatedVoluntary: Firm initiated