Z-2587-2016Class II

OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Ophthalmic: The OPTICAL BIOMETER AL-Scan is a...

Nidek Incorporated / initiated 2016-07-01 / Terminated / root cause: software

Nidek Inc began a recall of OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Ophthalmic: The OPTICAL BIOMETER AL-Scan is a measuring device that can measure as a single unit... on . FDA classified it as Class II and gives the reason as "Nidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2587-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nidek Incorporated
Product
OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Ophthalmic: The OPTICAL BIOMETER AL-Scan is a measuring device that can measure as a single unit the values necessary to calculate the power of an IOL for cataract surgery such as axial length, corneal curvature radius, and anterior chamber depth. It measures these necessary values successively through a non-contact optical measurement method.
Product code
HJO Biomicroscope, Slit-Lamp, Ac-Powered
Reason for recall
Nidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Dear Doctor/Recall letters were sent to affected customers on July 7, 2016. The letters describe the issue and provide instruction should the issue occur. The letters inform users that Nidek will reach out in the next few weeks to schedule installation of new software that will solve the problem. Questions or concerns may be directed to: Customer Service 1.800.722.0219; Monday- Friday 8 am to 5pm PST E-mail: [email protected].
Quantity in commerce
17
Distribution
US Nationwide Distribution in the states of: NY, PA. GA. AZ, VA, DE, ME,. MO, TN, and NM.
Product codes and lots
OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Serial numbers: 220005, 220006, 220008, 220009, 220013, 220017, 220018, 220019, 220020, 220021, 220023, 220025, 220026, 220031, 220032, 220033, 220036.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2587-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2587-2016Class IITerminatedVoluntary: Firm initiated