510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2250-2016Class II

Convenience Kits from Centurion that contain a Medtronic/Covidien Light Glove ED SUTURE KIT VAGINAL DELIVERY...

Centurion Medical Products, LP / initiated 2015-04-29 / Terminated / root cause: manufacturing process

Centurion Medical Products Corporation began a recall of Convenience Kits from Centurion that contain a Medtronic/Covidien Light Glove ED SUTURE KIT VAGINAL DELIVERY TRAY ENDOSCOPY TRAY MINOR BASIN SET LAPAROSCOPIC... on . FDA classified it as Class II and gives the reason as "Kits from Centurion contain a Medtronic/Covidien Light Glove that is being recalled by the manufacturer, Medtronic. According to the notice received from Medtronic, the recall was initiated because some light covers may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2250-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Centurion Medical Products, LP
Product
Convenience Kits from Centurion that contain a Medtronic/Covidien Light Glove ED SUTURE KIT VAGINAL DELIVERY TRAY ENDOSCOPY TRAY MINOR BASIN SET LAPAROSCOPIC BASIN TRAY
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
Kits from Centurion contain a Medtronic/Covidien Light Glove that is being recalled by the manufacturer, Medtronic. According to the notice received from Medtronic, the recall was initiated because some light covers may have splits or holes, which may result in the transfer of microorganisms from the light handle into the patient wound.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 4/28/2015, URGENT PRODUCT RECALL NOTICE notifications were sent to the affected customers via certified mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
Quantity in commerce
483
Distribution
KS, IN, IA, KY, MT, SD
Product codes and lots
Code Lot # Description exp date SUT18365 2012070350 ED SUTURE KIT Sep-14 MNS6710 2012070990 VAGINAL DELIVERY TRAY Jan-15 MNS6710 2012082090 VAGINAL DELIVERY TRAY Feb-15 MNS6710 2012092490 VAGINAL DELIVERY TRAY Apr-15 MNS6710 2012120390 VAGINAL DELIVERY TRAY Apr-15 MNS4635 2012081390 ENDOSCOPY TRAY Feb-15 OR1640 2012040990 MINOR BASIN SET Jan-15 OR1640 2012123190 MINOR BASIN SET Apr-15 OR1575 2012071690 LAPAROSCOPIC BASIN TRAY Nov-14 OR1575 2012091090 LAPAROSCOPIC BASIN TRAY Feb-15 OR1575 2013010490 LAPAROSCOPIC BASIN TRAY Apr-15

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2250-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2250-2016Class IITerminatedVoluntary: Firm initiated