510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2297-2016Class II

SPHERICAL HEAD REFLECTION(R) 6.5 MM CANCELLOUS SCREW, REF 71332525, QTY: (1), STERILE R, smith&nephew Total...

Smith & Nephew, Inc. / initiated 2016-07-06 / Terminated / root cause: packaging

Smith & Nephew, Inc. began a recall of SPHERICAL HEAD REFLECTION(R) 6.5 MM CANCELLOUS SCREW, REF 71332525, QTY: (1), STERILE R, smith&nephew Total hip components are indicated for individuals... on . FDA classified it as Class II and gives the reason as "The labels correctly state that the product should be a 25mm screw, but the box contained a longer, 35mm screw". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2297-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
SPHERICAL HEAD REFLECTION(R) 6.5 MM CANCELLOUS SCREW, REF 71332525, QTY: (1), STERILE R, smith&nephew Total hip components are indicated for individuals undergoing primary or revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant; congenital dysplasia; old, remote osteomyelitis with an extended drainage-free period, nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
The labels correctly state that the product should be a 25mm screw, but the box contained a longer, 35mm screw.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Smith & Nephew sent an Urgent Product Recall Notification letter dated July 6, 2016, to all affected consignees via email and FedEx Overnight delivery on July 6, 2016. The letter identified the product, the problem, and the action to be taken by the consignee. Customers were asked to inspect their inventory and located any unused affected devices and quarantine them immediately, complete the Inventory Return Certification Form and fax to 1-901-566-7975 indicating the quantities that need to be returned. Consignees should contact Smith & Nephew's Field Actions Department via email at [email protected] to obtain a return authorization (RA) number. For questions regarding this recall call 901-399-5520.
Quantity in commerce
185 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to AU, CA, GB, IN, ZA, and TR
Product codes and lots
Lot Number 16AM14274

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2297-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2297-2016Class IITerminatedVoluntary: Firm initiated