Z-2320-2016Class II
RePlant Angled Abutment Part Number 6050-52-60
Implant Direct Sybron Manufacturing LLC / initiated 2016-07-05 / Terminated / root cause: manufacturing process
Implant Direct Sybron Manufacturing, LLC began a recall of RePlant Angled Abutment Part Number on . FDA classified it as Class II and gives the reason as "Issue with design specification; oversized condition". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2320-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Implant Direct Sybron Manufacturing LLC
- Product
- RePlant Angled Abutment Part Number 6050-52-60
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- Issue with design specification; oversized condition.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Implant Direct sent an Urgent: Medical Device Field Corrective Action letter dated July 6, 2016, to all affected customers. The letter requests that it be reviewed and determined if customer has any affected product in inventory. Customers are asked to complete and return the Acknowledgement and Recall Return Form within 48 hours. If customers are an authorized Implant Direct Sybron Manufacturing distributor, they request that customers identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of the notification in order to provide the customers with the correct tool. If customers have any of the affected product listed the firm will send a replacement part. If any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425.
- Quantity in commerce
- 205
- Distribution
- U.S. distribution to the following; CA, WA, UT, GA, AZ, NE, OR, MI, TX, WI, FL, AL, AR, MO, VA, MA, TN, LA, NC, KS, NY, SC, OK. Foreign distribution to the following; CA, DE, IT, GB, PL, DE, GB, NL, GR, ES, HR, BG.
- Product codes and lots
- Lot #50174, 39799, 49364
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2320-2016 | Class II | Terminated | Voluntary: Firm initiated |